UF750TS

The Memmert pass-through oven saves time when loading and reduces the danger of contamination, especially when directly transporting the chamber load between the grey room and the clean room.

Category: Tag: Brand:

Temperature

Working temperature rangeat least 10 above ambient temperature to +250 °C
Setting accuracy temperatureup to 99.9 °C: 0.1 / from 100 °C: 0.5
Setting temperature range+20 to +250°C
resolution of display for actual values0.1°C
Temperature sensor2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error

Control technology

ControlCOCKPITTwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Language settingGerman, English, Spanish, French, Polish, Czech, Hungarian
TimerDigital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function HeatBALANCEadapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 %
Function SetpointWAITthe process time does not start until the set temperature is reached
Calibrationthree freely selectable temperature values

Ventilation

Fanforced air circulation by quite air turbine, adjustable in 10 % steps for each segment individually
Fresh airAdmixture of pre-heated fresh air by electronically adjustable air flap
Ventvent connection with restrictor flap

Communication

Documentationprogramme stored in case of power failure
ProgrammingAtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port

Safety

Temperature controlindependently working, digitally adjustable electronic micro-processor overtemperature monitor TWW, protection class 3.1 (max-value for overtemperature, min-value for undertemperature)
Temperature controlmechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature
AutoSAFETYadditionally integrated over- and undertemperature protection “ASF”, automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature
Autodiagnostic systemfor fault analysis
Alarmvisual and acoustic

Standard equipment

Works calibration certificateCalibration at +160°C
Doorfully insulated stainless steel doors on two sides
Internals2 stainless steel grid(s), electropolished
Installationwith feet

Stainless steel interior

Interioreasy-to-clean interior,made of stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides
Volume749 l
Dimensionsw(A) x h(B) x d(C): 1040 x 1200 x 600 mm
Max. number of internals14
Max. loading of chamber300 kg
Max. loading per internal30 kg

Textured stainless steel casing

Dimensionsw(D) x h(E) x d(F): 1224 x 1714 x 782 mm (d +2x56mm door handle)
pass-through version

Electrical data

Voltage Electrical load400 V and 3x 230 V w/o neutral, 50/60 Hz approx. 5000 W

Ambient conditions

Altitude of installationmax. 2,000 m above sea level
Ambient temperature+5 °C to +40 °C
Humidity rhmax. 80 %, non-condensing
Overvoltage categoryII
Pollution degree2

Packing/shipping data

Transport informationThe appliances must be transported upright
Customs tariff number8419 8998
Country of originFederal Republic of Germany
WEEE-Reg.-No.DE 66812464
Dimensions approx incl. cartonw x h x d: 1330 x 1920 x 1050 mm
Net weightapprox. 260 kg
Gross weight cartonapprox. 331 kg

Standard units are safety-approved and bear the test marks

 

Parameter

Temperature

Sector

Chemistry & Raw Material, Education & Research, Energy & Environment, Food & Beverage, Medical & Pharmacy, Metal & Electric

Size

Large (449 – 2200 liter volume)