ICO240

Safety at all times: High-end functions for the protection of cell cultures, bacteria cultures or tissue cultures.

Category: Tag: Brand:

Temperature

Working-temperature range5 °C above ambient temperature up to +50 °C
Standard sterilisation programme: 60 minutes at 180°C (without removing the sensors)
Setting temperature range+18 to +50 °C
Setting accuracy temperature0.1 °C
Temperature sensor2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Temperature variation in chamberat + 37 °C +/- 0.3 K
Temperature fluctuations with timeat 37 °C +/- 0.1 K

Humidity

Humidity control (standard)Humidity limitation thanks to a Peltier element; when water dish is full and inserted, the Peltier element limits the value of relative humidity in the interior to 93 % rh +/- 2.5 %
Setting accuracy humidity0.5 % rh
Setting range active humidity control (with option K7)40 to 97 % rh and rh-Off

Control of standard components

CO2 controlDigital electronic CO2 control with dual beam NDIR system, with auto-diagnostic system and acoustic fault indication, barometric pressure compensation
Adjustment range CO20 to 20 % CO2
Setting accuracy CO20,1%
Variation in time CO2+/- 0.2 % CO2
Adjustment range O21 to 20 % O2
Setting accuracy O20.1 % O2

Control technology

ControlCOCKPITTwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Language settingGerman, English, Spanish, French, Polish, Czech, Hungarian
Function SetpointWAITthe process time does not start until the set temperature is reached
Adjustable parameterstemperature (Celsius or Fahrenheit), CO2, programme time, time zones, summertime/wintertime

Communication

InterfaceEthernet LAN, USB
Documentationprogramme stored in case of power failure
ProgrammingAtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port

Safety

AutoSAFETYadditionally integrated over- and undertemperature protection “ASF”, automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating function is switched off in case of overtemperature, cooling function in case of undertemperature
Autodiagnostic systemintegral fault diagnostics for temperature and CO2
Alarmvisual and acoustic

Heating concept

6 sideslarge-area multi-function heating system on four sides with additional door and back heating to avoid condensation

Standard equipment

Works calibration certificateWorks calibration certificate at +37 °C, 5 % CO2 and 90 % rh (requires option K7); standard equipment for units with active humidity control
Works calibration certificateWorks calibration certificate at +37 °C, 5 % CO2, 90 % rh and 10 % O2 (requires option K7 and option T6); standard equipment for units with O2 control
Works calibration certificateincl. works calibration certificate (measuring point chamber centre) at +37°C, 5 % CO2 for standard units
Doorfully insulated stainless steel door with 2-point locking (compression door lock)
Doorinner glass door with opening (8 mm Ø) to take gas sample
Standard equipmentCO2 connection set: hose with coupling and clamp
Standard accessoriesMembrane filter (in order to remove impurities and pollutants, all incoming gases pass through a membrane filter before they reach the chamber)
Internals2 perforated stainless steel shelf/shelves
Internals1 stainless steel water dish

Stainless steel interior

Dimensionsw(A) x h(B) x d(C): 600 x 810 x 500 mm (d less 35 mm for fan)
Interiormaterial 1.4301 (ASTM 304), corrosion resistant
Volume241 l
Max. number of internals12
Max. loading of chamber140 kg
Max. loading per internal15 kg

Textured stainless steel casing

Dimensionsw(D) x h(E) x d(F): 759 x 1180 x 691 mm (d +56mm door handle)
Housingrear zinc-plated steel

Electrical data

Voltage Electrical load230 V, 50/60 Hz
approx. 1650 W
Voltage Electrical load115 V, 50/60 Hz
approx. 1650 W

Ambient conditions

Set UpThe distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Ambient temperature10 °C to 35 °C
Humidity rhmax. 70 %, non-condensing
Altitude of installationmax. 2,000 m above sea level
Overvoltage categoryII
Pollution degree2

Packing/shipping data

Transport informationThe appliances must be transported upright
Customs tariff number8419 8998
Country of originFederal Republic of Germany
WEEE-Reg.-No.DE 66812464
Dimensions approx incl. cartonw x h x d: 840 x 1360 x 900 mm
Net weightapprox. 110 kg
Gross weight cartonapprox. 145 kg

Standard units are safety-approved and bear the test marks