SN75plus

Sterilisation instead of disinfection of instruments in the laboratory. The steriliser complies with all relevant standards and regulations for medical devices.

Category: Tags: ,

Temperature

Setting temperature range+20 to +250 °C
Setting accuracy temperatureup to 99.9 °C: 0.1 / from 100 °C: 0.5
Working temperature rangeat least 5 above ambient temperature to +250 °C
Temperature sensor2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error

Control technology

ControlCOCKPITTwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
TimerDigital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function HeatBALANCEadapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 %
Function SetpointWAITthe process time does not start until the set temperature is reached
Calibrationthree freely selectable temperature values
adjustable parameterstemperature (Celsius or Fahrenheit), air flap position, programme time, time zones, summertime/wintertime

Ventilation

Convectionnatural convection
Fresh airAdmixture of pre-heated fresh air by electronically adjustable air flap
Ventvent connection with restrictor flap

Communication

Documentationprogramme stored in case of power failure
ProgrammingAtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port

Safety

Temperature controlmechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature
Temperature controlovertemperature monitor TWW, protection class 3.1 or adjustable temperature limiter TWB, protection class 2, selectable on display
AutoSAFETYadditionally integrated over- and undertemperature monitor “ASF”, automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature
Autodiagnostic systemfor fault analysis
Alarmvisual and acoustic

Standard equipment

Works calibration certificateCalibration at +160°C
Doorfully insulated stainless steel door with 2-point locking (compression door lock)
Internals2 stainless steel grid(s), electropolished

Stainless steel interior

Interioreasy-to-clean interior,made of stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides
Volume74 l
Dimensionsw(A) x h(B) x d(C): 400 x 560 x 330 mm
Max. number of internals6
Max. loading of chamber120 kg
Max. loading per internal20 kg

Textured stainless steel casing

Dimensionsw(D) x h(E) x d(F): 585 x 944 x 514 mm (d +56mm door handle)
Housingrear zinc-plated steel

Electrical data

Voltage Electrical load230 V, 50/60 Hz
approx. 2500 W
Voltage Electrical load115 V, 50/60 Hz
approx. 1800 W

Ambient conditions

Set UpThe distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Altitude of installationmax. 2,000 m above sea level
Ambient temperature+5 °C to +40 °C
Humidity rhmax. 80 %, non-condensing
Overvoltage categoryII
Pollution degree2

Packing/shipping data

Transport informationThe appliances must be transported upright
Customs tariff number8419 8998
Country of originFederal Republic of Germany
WEEE-Reg.-No.DE 66812464
Dimensions approx incl. cartonw x h x d: 730 x 1130 x 670 mm
Net weightapprox. 66 kg
Gross weight cartonapprox. 85 kg

Standard units are safety-approved and bear the test marks

 

 

 

Intended use as medical device

Memmert is bringing medical devices of class I according to EU Directive (EU) 2017/745 into circulation.

TypeIntended use as medical deviceClassification according to (EU) 2017/745
UNmplus, UFmplusThe appliance may be used for heating fango, silicate and APS packs for physical therapy and keeping them warm.I
UNm, UFm, INm, IFmThe appliance may be used for heating fango, silicate and APS packs for physical therapy and keeping them warm.I
INmplus, IFmplusIm may be used for heating fango, silicate and APS packs for physical therapy and keeping them warm. Implus may be used for temperature control of rinsing and infusion solutions and contrast agents.I
IFbwThe appliance is used to heat non-sterile cloths and covers. Any other use is improper
and may cause damage or danger.
I

Memmert is bringing medical devices of class IIa and IIb into circulation according to MDD 93/42/EEC until 31.12.2028 according to the transitional provisions (EU) 2023/607 by the (EU) 2017/745 article 120 (2).

TypeIntended use as medical deviceClassification according to 93/42/EWG
SNplus, SFplusThe product is intended for the sterilisation of medical devices with dry heat at atmospheric pressure.IIb
SN, SFThe product is intended for the sterilisation of medical devices with dry heat at atmospheric pressure.IIb
ICOmedThe CO2 incubator ICOmed is used to generate and maintain constant ambient conditions for the in-vitro fertilisation (IVF) application field, especially for the incubation of oocytes, spermatozoa and zygotes in special culture dishes for IVF application as well as for gene expression and the biosynthesis of RNA and proteins.IIa

 

Air Circulation

Natural air circulation (Convection)