Precise drying, heating, ageing, burn-in and hardening in research, science, industry and quality assurance.

Temperature

Working temperature rangeat least 5 (UN/UNplus/UNm/UNmplus) or 10 (UF/UFplus/UFm/UFmplus) above ambient temperature to +300 °C
Setting accuracy temperatureup to 99.9 °C: 0.1 / from 100 °C: 0.5
Setting temperature range+20 to +300 °C
Temperature sensor2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error

Control technology

ControlCOCKPITTwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
TimerDigital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function HeatBALANCEadapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 %
Function SetpointWAITthe process time does not start until the set temperature is reached
Calibrationthree freely selectable temperature values
adjustable parameterstemperature (Celsius or Fahrenheit), fan speed, air flap position, programme time, time zones, summertime/wintertime

Ventilation

Fanforced air circulation by quite air turbine, adjustable in 10 % steps for each segment individually
Fresh airAdmixture of pre-heated fresh air by electronically adjustable air flap
Ventvent connection with restrictor flap

Communication

Documentationprogramme stored in case of power failure
ProgrammingAtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port

Safety

Temperature controlmechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature
Temperature controlovertemperature monitor TWW, protection class 3.1 or adjustable temperature limiter TWB, protection class 2, selectable on display
AutoSAFETYadditionally integrated over- and undertemperature monitor “ASF”, automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature
Autodiagnostic systemfor fault analysis
Alarmvisual and acoustic

Standard equipment

Works calibration certificateCalibration at +160°C
Doorfully insulated stainless steel door with 2-point locking (compression door lock)
Internals2 stainless steel grid(s), electropolished

Stainless steel interior

Dimensionsw(A) x h(B) x d(C): 560 x 720 x 400 mm (d less 39 mm for fan)
Interioreasy-to-clean interior,made of stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides
Volume161 l
Max. number of internals8
Max. loading of chamber210 kg
Max. loading per internal20 kg

Textured stainless steel casing

Dimensionsw(D) x h(E) x d(F): 745 x 1104 x 584 mm (d +56mm door handle)
Housingrear zinc-plated steel

Electrical data

Voltage Electrical load230 V, 50/60 Hz
approx. 3200 W
Voltage Electrical load115 V, 50/60 Hz
approx. 1800 W

Ambient conditions

Set UpThe distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Altitude of installationmax. 2,000 m above sea level
Ambient temperature+5 °C to +40 °C
Humidity rhmax. 80 %, non-condensing
Overvoltage categoryII
Pollution degree2

Packing/shipping data

Transport informationThe appliances must be transported upright
Customs tariff number8419 8998
Country of originFederal Republic of Germany
WEEE-Reg.-No.DE 66812464
Dimensions approx incl. cartonw x h x d: 830 x 1300 x 800 mm
Net weightapprox. 96 kg
Gross weight cartonapprox. 122 kg