WTB6

Ideal for the temperature control of ointments, emulsions, samples, plates and nutrient solutions in the laboratory, as well as warm storage and calibration in the industrial segment

Category: Tag: Brand:

Temperature

Temperature sensordigital temperature sensor, mounted on the outside of the bath
TemperatureSetting Celsius / Fahrenheit
Working temperature rangeat least 5 above ambient to +100 (with cover, depending on ambient pressure)
Setting temperature range+10 to +100 °C
Setting accuracy temperature0.1 °C

Control technology

Calibration2-point calibration
Controllerintuitive 3.5″ colour graphic display with touch screen
Controllerdigital display of set and actual temperature and of (remaining) programme time

Safety

Temperature controltwo-stage safety system to prevent overtemperature (pressure sensor for level monitoring, thermal fuse for heating shutdown in case of error)
Temperature controlindividually adjustable overtemperature alarm, visual and acoustic
Temperature controlIndependent temperature limiter to switch off the heating in case of failure
Autodiagnostic systemmicroprocessor PID-temperature controller with integrated autodiagnostic system with fault indicator

Heating concept

Heatingcorrosion-proof large-area heating on the outside of the interior

Standard equipment

CoverStainless steel gable cover
Drain systeminnovative drain system – easy to empty
Timerdigital timer from 1 min. up to 99 hours 59 min., visual and acoustic

Stainless steel interior

InteriorDeep-drawn stainless steel tray (stainless steel W.St. 1-4301), without corners and edges for easy cleaning incl. central drain (no disturbing installations)
Dimensionsl(A) x w(B) x h(C): 277 x 217 x 150 mm
Volume7.5 l
Liquid level min.40 mm
Liquid level max.114 mm

Textured stainless steel casing

Dimensionsl(D) x w(E) x h(F2): 479 x 289 x 375 mm

Electrical data

Voltage Electrical load230 V, 50/60 Hz
approx. 1000 W
Voltage Electrical load115 V, 50/60 Hz
approx. 1000 W

Ambient conditions

InstallationThe vent openings in the left and back side must remain unobstructed. The distance to the wall must be at least 100 mm on the sides and at least 150 mm to the rear. The minimum spacing from the top of the bath to the next ceiling is 750 mm.
Ambient temperature+5 °C to +40 °C
Humidity rhmax. 80 %, non-condensing
Overvoltage categoryII
Pollution degree2
Water bath

Packing/shipping data

Dimensions approx incl. cartonl x w x h: 579 x 389 x 475 mm
Transport informationThe appliances must be transported upright
Customs tariff number8419 8998
Country of originFederal Republic of Germany
WEEE-Reg.-No.DE 66812464
Net weightapprox. 11 kg
Gross weight cartonapprox. 13 kg

Standard units are safety-approved and bear the test marks