Temperature
Working temperature range |
at least 10 above ambient temperature to +250 °C |
Setting accuracy temperature |
up to 99.9 °C: 0.1 / from 100 °C: 0.5 |
Setting temperature range |
+20 to +250°C |
resolution of display for actual values |
0.1°C |
Temperature sensor |
2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error |
Control technology
ControlCOCKPIT |
TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays. |
Language setting |
German, English, Spanish, French, Polish, Czech, Hungarian |
Timer |
Digital backwards counter with target time setting, adjustable from 1 minute to 99 days |
Function HeatBALANCE |
adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 % |
Function SetpointWAIT |
the process time does not start until the set temperature is reached |
Calibration |
three freely selectable temperature values |
Ventilation
Fan |
forced air circulation by quite air turbine, adjustable in 10 % steps for each segment individually |
Fresh air |
Admixture of pre-heated fresh air by electronically adjustable air flap |
Vent |
vent connection with restrictor flap |
Communication
Documentation |
programme stored in case of power failure |
Programming |
AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port |
Safety
Temperature control |
independently working, digitally adjustable electronic micro-processor overtemperature monitor TWW, protection class 3.1 (max-value for overtemperature, min-value for undertemperature) |
Temperature control |
mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature |
AutoSAFETY |
additionally integrated over- and undertemperature protection “ASF”, automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature |
Autodiagnostic system |
for fault analysis |
Alarm |
visual and acoustic |
Standard equipment
Works calibration certificate |
Calibration at +160°C |
Door |
fully insulated stainless steel doors on two sides |
Internals |
2 stainless steel grid(s), electropolished |
Installation |
with feet |
Stainless steel interior
Interior |
easy-to-clean interior,made of stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides |
Volume |
161 l |
Dimensions |
w(A) x h(B) x d(C): 560 x 720 x 400 mm |
Max. number of internals |
8 |
Max. loading of chamber |
210 kg |
Max. loading per internal |
20 kg |
Textured stainless steel casing
Dimensions |
w(D) x h(E) x d(F): 745 x 1233 x 582 mm (d +2x56mm door handle) |
pass-through version |
Electrical data
Voltage Electrical load |
230 V, 50/60 Hz
approx. 3200 W |
Voltage Electrical load |
115 V, 50/60 Hz
approx. 1800 W |
Ambient conditions
Altitude of installation |
max. 2,000 m above sea level |
Ambient temperature |
+5 °C to +40 °C |
Humidity rh |
max. 80 %, non-condensing |
Overvoltage category |
II |
Pollution degree |
2 |
Packing/shipping data
Transport information |
The appliances must be transported upright |
Customs tariff number |
8419 8998 |
Country of origin |
Federal Republic of Germany |
WEEE-Reg.-No. |
DE 66812464 |
Dimensions approx incl. carton |
w x h x d: 830 x 1300 x 800 mm |
Net weight |
approx. 120 kg |
Gross weight carton |
approx. 146 kg |
Standard units are safety-approved and bear the test marks